CVE-2025-14097
Fix Status by Distribution
Fixed in 0 of 27 tracked Linux releases
Fix evidence last checked: 2026-07-24 08:46 UTC
Get alerted when your releases are fixedNo fix evidence found:
A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software. Other related CVE's are CVE-2025-14095 & CVE-2025-14096. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer. Temporary work Around: If the network is not considered secure, please remove the analyzer from the network. Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.
Exploit evidence
Last checked: 2026-07-31 19:14 UTC
No exploitation evidence · CISA Coordinator (Assessed 2025-12-17 14:40 UTC)